Aquinity² P35 / P70 Ultra Pure Water System
- ✓ Compact housing
- ✓ Big touchscreen
- ✓ Remote service module
- ✓ Menus and wizards easy to use and mistake avoiding
- ✓ Fixed and flexible dispenser
- ✓ Easy access for cartridge exchange
Designed for the production of pure water and ultra pure water out of tap water
- Aquinity² P versions produce deionised water (< 1 µS/cm / > 1 MOhm x cm) by reverse osmosis and conditioning module into the external reservoir (35 liter or 70 liter).
- The water can be tapped directly (e.g. for glassware cleaning by hand) or connecting the tank to a laboratory dishwasher.
- Out of the storage tank the ultra pure water is produced and then can be tapped with a flexible dispenser.
- The Aquinity² water purification system is available in different configurations (Reagent, Analytical & LifeScience) to meet the specific requirements of ultra pure water for different applications.
- Analytical is a version with UV lamp for low TOC value. Life Science version comes with UV lamp and Ultrafiltration Module for low TOC and low endotoxin values.
- An optional integrated TOC sensor allows the user to monitor the TOC content permanently online.
- Besides that the Aquinity² can be configured as bench top or bench-integrated system with external panel and wall mounted (with inclined display).
- The tank can be installed at the side or above the system or even underneath the bench.
- Upon request the tanks can be connected in series to reach a bigger volume of reservoir.
Main technical specifications of the system
18.2 MOhm x cm | 0.055 µS/cm; Type I
< 5 ppb (Life Science)
< 3 ppb (Analytical)
System flow scheme
The scheme shows an Aquinity P35/70 Life Science with integrated TOC-monitoring.

Flexible to fit your lab

Bench integrated configuration (BI version)
To save valuable benchtop space the system can be ordered as BI configuration. The system is installed underneath the work bench and only the display and dispenser are mounted on the wall.

Wall mounted configuration
To save valuable benchtop space the system can be ordered as wall mounted configuration. The whole system can be mounted on the wall with inclined display.

External pump
Compact pump is required when feeding laboratory equipment such as autoclaves or laboratory dish washers.

TOC-Monitoring (TI version)
The TOC monitoring during production and intermittent measurements during non-use periods allows checking the organic content in water continuously. The TOC value is measured between 1 and 999 ppb and shown on the display.
Specifications & Requirements
| Pure water + ultrapure water quality | < 2 µS/cm; Type II + 0.055 µS/cm; Type I |
|---|---|
| Resistivity | > 0.5 MegaOhm; Type II + 18.2 MegaOhm/cm; Type I |
| Total organic carbon (TOC) | < 10 ppb Reagent, < 5 ppb Life Science, < 3 ppb Analytical |
| Dispensing flow rate | 2.0 L/min (1.5 L/min Life Science) |
| Production rate | 10 L/h, optionally 20 L/h |
| Bacteria | < 1 cfu/mL |
| Particulate | > 0.2 µm less than 1 particulate/mL |
| Pyrogen (endotoxins) | < 0.001 EU/mL* |
| RNAse | < 1 pg/mL* |
| DNAse | < 5 pg/mL* |
| Dimensions, weight, power | 504 x 680 x 535 mm, 16–20 kg, 110 – 230 V |
| *Life Science | |
| Type of feed water | tap water |
|---|---|
| Feed water conductivity | < 1400 µS/cm |
| Inlet pressure | 1.5 to 6 bar |
| Free chlorine | < 0.1 mg/L |
| Silt Density Index (SDI) | < 3 |
| pH | 3 to 9 |
| Temperature | 5 to 25 °C |
Get a quote now