flexiTOC Analyzer
The next-generation TOC analyzer for tap water to demanding samples. UV oxidation of organic carbon compounds to CO₂, quantified by a highly sensitive NDIR detector for excellent detection limits and measurement stability.
- ✓ Wide range — tap water to demanding samples
- ✓ UV oxidation with sensitive NDIR CO₂ detection
- ✓ Reagent-free operation below 2000 ppb TOC
- ✓ Optimized UV reactor geometry & harmonized gas flow
- ✓ Cloud data export & remote diagnosis
- ✓ Up to 4 inlets — multiline capable
Wide-range TOC with sensitive NDIR detection
UV oxidation of organic carbon compounds in aqueous samples to carbon dioxide (CO₂). The generated CO₂ is quantified using a highly sensitive NDIR detector, ensuring excellent detection limits and measurement stability. The system supports reagent-free operation for TOC levels below 2000 ppb, and operation with reagent for higher TOC concentrations and less-oxidative substances, depending on the application.
Configure the flexiTOC for your process
Multiline mode
The flexiTOC pro provides 4 inlets; a software upgrade lets you measure up to 3 loops sequentially, configuring each interval separately.
Sample mode
A sample-mode configuration measures separate samples offline; the sequence is user-customizable.
Calibration & verification
Reagent and water-blank determination plus multipoint calibration with internal standard dilution for enhanced analytical accuracy.
Data export
Export to external storage in various formats; cloud storage via network, and sharing with customer service for error diagnosis.
System status
A complete software overhaul delivers a detailed status overview, including operating hours for individual sensors, a pump, and the UV lamp.
Heat exchanger
Required for medium temperatures above 50 °C.
Pressure regulator
A pressure transducer is required if the incoming medium pressure exceeds 0.5 bar.
21 CFR Part 11 software
Conformity for pharma applications — secure data handling, electronic records and signatures.
IP54 housing
Screen and fan supplied with IP54 protection as standard.
Reagent-free & reagent modes
The flexiTOC runs reagent-free for TOC levels below 2000 ppb, and switches to reagent operation for higher TOC concentrations and less-oxidative substances — covering a wide application range from tap water to demanding samples.
Main technical specifications
| Method & detection | UV oxidation with CO₂ detection (NDIR) |
|---|---|
| Type of sample | tap to ultra pure water |
| Sample volume | 5 – 30 mL reagent or dilution water is added automatically |
| Period of measurement | Sample interval 5 –10 min, user defined interruptions up to 24h |
| Measurement range | 0.05 to 20 ppm, up to 100 ppm in diluting mode |
| Detection accuracy | < 100 ppb ± 50 ppb; 100… 1000 ppb max. 10 %; above 1000 ppb 2.5 % |
| Ports | 1 for sample, 1 for dilution water, 1 for reagent | 1 for sample, 3 for SST |
| Related norms | USP <643>, EP 2.2.44, DIN EN 1484 |
| Ambient parameters | relative humidity: 10 to 95 %, temperature: 10 to 45 °C |
| Interface ports | analog (0 – 20 mA), according to Namur NE43 |
| Carrier gas | N2, O2, compressed air, gas pressure: 1 – 4 bar, consumption: approx. 12 L/h |
| Inlet pressure & temperature | < 0.5 bar (> 0.5 bar pressure reducer is needed) & 50 °C |