flexiTOC Analyzer
The next-generation TOC analyzer for tap water to demanding samples. UV oxidation of organic carbon compounds to CO₂, quantified by a highly sensitive NDIR detector for excellent detection limits and measurement stability.
Overview
UV oxidation of organic carbon compounds in aqueous samples to carbon dioxide (CO₂). The generated CO₂ is quantified using a highly sensitive NDIR detector, ensuring excellent detection limits and measurement stability.
The system supports reagent-free operation for TOC levels below 2000 ppb, and operation with reagent for higher TOC concentrations and less-oxidative substances, depending on the application.
Highlights
- ✓ Wide range: tap water to demanding samples
- ✓ UV oxidation with sensitive NDIR CO₂ detection
- ✓ Reagent-free operation below 2000 ppb TOC
- ✓ Optimized UV reactor geometry & harmonized gas flow
- ✓ Cloud data export & remote diagnosis
Options & Features
Multiline mode
The flexiTOC pro provides 4 inlets; a software upgrade lets you measure up to 3 loops sequentially, configuring the duration of each interval measurement separately.
Sample mode
A sample-mode configuration measures separate samples offline; the sequence is user-customizable.
Calibration & verification
Reagent and water-blank determination plus multipoint calibration with internal standard dilution for enhanced analytical accuracy.
Data export
Export to an external storage device in various formats; cloud storage via network, and sharing with customer service for error diagnosis.
System status
A complete software overhaul delivers a detailed status overview, including operating hours for individual sensors, a pump, and the UV lamp.
Heat exchanger & pressure regulator
A heat exchanger is required for medium temperatures above 50 °C; a pressure transducer is required if the incoming medium pressure exceeds 0.5 bar.
21 CFR Part 11 software · IP54
Conformity for pharma applications (secure data handling, electronic records and signatures). Screen and fan supplied with IP54 protection as standard.
Specifications & Requirements
Method & sample
Method & detection: UV oxidation with CO₂ detection (NDIR)
Type of sample: tap water to demanding samples
Sample volume: 5–30 mL reagent or dilution water added automatically
Period of measurement: sample interval 5–10 min; user-defined interruptions up to 24 h
Measurement range: 0.05 to 20 ppm, up to 100 ppm in diluting mode
Detection accuracy: < 500 ppb ± 5 ppb; > 500 ppb ± 1% (non-diluting mode)
Ports & environment
Ports: 1 sample, 1 dilution water, 1 reagent | 1 sample, 3 for SST
Related norms: USP <643>, EP 2.2.44, DIN EN 1484
Ambient parameters: relative humidity 10–95%; temperature 10–45 °C
Interface ports: analog (0–20 mA), according to NAMUR NE43
Carrier gas: N₂, O₂, compressed air; 1–4 bar; consumption approx. 12 L/h
Inlet pressure & temperature: < 0.5 bar (> 0.5 bar pressure reducer needed) & 50 °C